CHAPTER 3 Settings and Measurement
(2) DATA settings
This menu is used to specify capture-related items and calculations.
Sampling
Capture destination
Ring Capture
External Sampling
AC Line Filter
Backup
Calc. Settings 1
Calc. Settings 2
Captured data file name
Folder (File)
Name Type
File Type
(2)-1 Sampling interval
Sets the sampling interval for data capture.
The table below shows the number of measuring channels and sampling interval values that can be set.
If data fluctuate due to noise, set the sampling interval to a value which enables the digital filter function.
Number of Measuring Channels
*1: "Number of Measuring Channels" is the number of channels in which input settings are NOT set to "OFF".
*2: The temperature setting is not available for sampling intervals of 10, 20, and 50 ms
To use the digital filter function, you must set the AC power supply frequency accurately. Follow the instructions on page 3-28
to ensure that the settings are accurate.
3−24
Setting
File Name
[ON]
Number of
Ring Capture
Points
Backup interval
Backup destination
Save folder
Setting
Capture destination : MEM, USB1
Folder
File
Auto, Arbitrary, Sequential number
Binary (GBD), Text (CSV)
1 CH
2 CH
3 CH to 5 CH
6 CH to 10 CH
11CH to 20CH
21CH to 50CH
51CH to 100CH
101CH to 200CH
10, 20, 50, 100, 125, 200, 250, 500 ms; 1, 2, 5, 10, 20, 30s; 1, 2, 5, 10, 20,
30 min; 1h
* Available sampling intervals vary depending on the input settings and the
number of channels to be used.
* Refer to "(2)-1 Sampling Interval" for details.
Internal memory, USB memory
* See the description of a captured data file name given below.
Off, On
1000 to 2000000 (A warning message will appear if you set a value larger
than the remaining disk space.)
Off, On
Off, On
Off, 1, 2, 6, 12, 24 hours
USB1, FTP
Folder name
Off, Average, Max, Min, Peak, RMS
Off, Average, Max, Min, Peak, RMS
Selections available
: Text input (if the naming method is Auto)
: Text input (if the naming method is Arbitrary or Sequential number)
Allowed Sampling Interval
*1
10 ms or slower *2
20 ms or slower *2
50 ms or slower *2
100 ms or slower
200ms or slower
500ms or slower
1s or slower
2s or slower
Selections available
Sampling Interval which enables
Digital Filter
50 ms or slower
125 ms or slower
250 ms or slower
500 ms or slower
1s or slower
2s or slower
5s or slower
10s or slower