Troubleshooting During Startup - Keysight E5061B Installation Manual

Network analyzer
Hide thumbs Also See for E5061B:
Table of Contents

Advertisement

2
Troubleshooting

Troubleshooting during Startup

50
When you encounter problems during start up, see
System recovery saves most of problems.
Table 3
Troubleshooting during startup
Symptom
Turning on (|) the standby switch does
not start up the system.
The system starts up, but it automatically
shuts down immediately.
The system starts up, but it enters the
service mode (The status bar at the lower
right of the screen displays SVC in red).
The measurement screen appears after
start up, but the power-on test fails
with Error Message 241 prompting on a
red background in the instrument's
message/warning area at the lower left
of the screen.
Table
Solution
Confirm that the power cable is
properly plugged in.
Confirm that the line switch at the
rear panel is turned on.
Execute the system recovery.
Network Analyzer
3.

Advertisement

Table of Contents
loading

Table of Contents