...5.6 Configure Outputs

7. 0 0
pH
0. 0
pH
AO1: Zero Value
Hold
On
-----
Off
mA
AO1: Default
Config. Output 1
Off
or
Hold
On
12. 0 0
mA
AO1: Default Val
Config. Output 1
Zero Value
(or
or
or
pH
mV
Deg.C
Deg.F
line. Use the
and
keys to adjust the displayed reading to the required zero value:
Note. The zero value setting plus the minimum differential determines the minimum and
maximum values for the span setting, e.g. to set a span value of –1100mV, first set the zero
value to –1200mV.
pH
– –2.00 to 14.00 pH (minimum differential, 2.00pH)
– –1200 to 1100 mV (minimum differential, 100mV)
ORP/Redox
Temperature
– Deg. C –10.0 to 140.0 (minimum differential, 10C)
Deg. F 14.0 to 284.0 (minimum differential, 18F)
A-B
– –2.00 to 12.00 pH (minimum differential, 2.00pH)
Note.
is applicable only to dual input analyzers.
A-B
Default Output
Select the system reaction to failure:
– Hold the analog output at the value prior to the failure.
Hold
On
– Stop on failure. This drives the analog output to the level set in the
frame below.
Off
– Ignore failure and continue operation.
Default Value
The level to which the analog output is driven if a failure occurs.
Set the value between 0.00 and 22.00mA
Output 2 configuration (and Outputs 3 and 4 if option board fitted and analog
Config. Output 2
features enabled – see Section 7.3) is identical to Output 1 configuration.
Option board fitted and analog features enabled (Section 7.3) – see Section 5.7.
CONFIG. CLOCK
Option board fitted and Serial Communications feature enabled (Section 7.3) –
CONFIG. SERIAL
see Supplementary Manual PROFIBUS Datalink Description (IM/PROBUS).
Single input analyzer  option board not fitted – see PID Control
CONFIG. CONTROL
Supplement (IM/AX4PID).
Dual input analyzer  option board not fitted – see Section 5.8.
CONFIG. SECURITY
or
) and
are shown alternately on the center display
A-B
Adjust
5 PROGRAMMING...
Default Val
39