...7.3 Factory Settings
Option Board
Fitted and Analog
Features Enabled
-----
Cal. Output 3
-----
Cal. Output 4
Option Board Fitted,
Additional Features
Disabled
Analog
-----
Pb Dp
Option Board
Option Board
Not Fitted
0000
Alter Fact. Code
Calibrate Output 3
Notes.
• Output 3 (and Output 4) calibration is applicable only if the Option board is fitted and
analog features enabled – see below.
• Output 3 calibration is identical to Output 2 calibration.
Calibrate Output 4
Note. Output 4 calibration is identical to Output 3 calibration.
Configure Option Board
Notes.
• This frame is displayed only if an option board is fitted.
• The software detects if an option board is fitted but cannot detect the additional features
available.
• If an option board is fitted, the correct selection must be made below to enable use of the
available features. If an incorrect selection is made, the software menus and frames
associated with that option are displayed in the Operating and Configuration pages but
the features do not work.
Use the
and
keys to enable the features for the type of option(s) fitted:
Analog
– Analog features enabled (comprising two additional analog
outputs, two additional alarm relays, clock and logbook facility).
Pb Dp
– PROFIBUS-DP digital communications features enabled.
Analog + Pb Dp
– Both analog and PROFIBUS-DP features enabled.
Alter Factory Code
Set the factory settings access code to a value between 0000 and 19999.
Return to main menu.
FACTORY SETTINGS
Monitoring pH
Redox (ORP)
Operating Page – see Section 2.3.
Dual Redox
pH/Redox (ORP)
7 CALIBRATION