...5 PROGRAMMING
...5.5 Configure Alarms

5.5.1 Wash Cycle Configuration (applicable only to Alarm 3)

A3: Type
set to
Wash
Cont.
-----
Pulsed
Wash Mode
Hours
15
Mins
Wash Frequency
Mins
15
Secs
Wash Duration
1. 0
Mins
Recovery Period
Config. Alarm 3
Continuous
Pulsed
36
Wash Mode
Select the wash mode required.
Cont.
– (continuous) the relay remains energized for the wash duration
– the relay is switched on and off every second for the duration of the wash,
Pulsed
– see Fig. 5.1
Wash Frequency
Set the wash frequency required.
Wash frequency is set in 15 minute increments between 15 and 45 minutes, then in 1 hour
increments between 1 and 24 hours.
Wash Duration
Set the wash duration required.
Wash duration is set in 15 second increments between 15 and 45 seconds, then in 1
minute increments between 1 and 10 minutes.
Recovery Period
Set the recovery period required, between 0.5 and 5.0 minutes in 0.5 minute increments.
Option board fitted and analog features enabled (Section 7.3) – Alarm 4
Config. Alarm 4
configuation is identical to Alarm 1.
Option board not fitted or option board fitted and analog features disabled
CONFIG. OUTPUTS
(Section 7.3) – see Section 5.6.
Wash Duration
1s 1s
Fig. 5.1 Pulsed and Continuous Wash Cycles
Frequency
Recovery Period
t