...5 PROGRAMMING
5.6 Configure Outputs
-----
CONFIG. OUTPUTS
-----
Config. Output 1
A-B
Temp.B
Sen.B
Temp.A
-----
Sen.A
AO1: Assign
4-20mA
0-20mA
-----
0-10mA
AO1: Range
7. 0 0
pH
0. 0 0
pH
AO1: Span Value
38
Configure Output 1
Output 2 configuration (and Outputs 3 and 4 if option board fitted and analog
Config. Output 2
features enabled – see Section 7.3) is identical to Output 1 configuration.
Assign
Select the sensor and analog output required:
Sen.A
– pH/Redox (ORP)/mV for the selected sensor.
Sen.B
Temp.A
– Temperature for the selected sensor.
Temp.B
A-B
Difference between the Sensor A and Sensor B readings.
Note.
,
and
Sen.B
Temp.B
A-B
Range
Select the analog output current range for the selected output.
Span Value
(or
or
or
pH
mV
Deg.C
Deg.F
line. Use the
and
keys to adjust the displayed reading to the required span value.
Note. The minimum and maximum span values are determined by the
(see next page) plus the minimum differential, e.g. to set
to –2.00pH.
Zero Value
pH
– 0.00 to 16.00 pH (minimum differential, 2.00pH)
Redox/ORP
– –1100 to 1200 mV (minimum differential, 100mV)
– Deg. C 0.0 to 150.0 (minimum differential, 10C)
Temperature
Deg. F 32.0 to 302.0 (minimum differential,18F)
A-B
– 0.00 to 14.00 pH (minimum differential, 2.00pH)
Note.
is applicable only to dual input analyzers.
A-B
AO1: Zero Value
Continued on next page.
are applicable only to dual input analyzers.
or
) and
are shown alternately on the upper display
A-B
Adjust
setting
Zero Value
to 0.00 pH, first set
Span Value