Manual, Single- And Two-Point Calibration (Ph Only) - ABB AX416 User Manual

Single and dual input analyzers for ph/redox (orp)
Hide thumbs Also See for AX416:
Table of Contents

Advertisement

...4 SETUP
...4.1 Sensor Calibration

...4.1.5 Manual, Single- and Two-Point Calibration (pH Only)

A: Buffer Method
set to
Man 2-Pt
9. 1 8
pH
20. 0
Deg.C
A: Immerse Buf.2
9. 2 2
pH
105
mV
Press
button
when stable
9. 2 2
pH
A: Cal. Accepted
9. 2 2
pH
A: Cal. Buffer 2
A: Buffer Method
set to
Man 1-Pt
-----
Passed
A: Calibration
100. 0
%
7. 0 0
pH
A: Slope & Check
Sensor Cal. A
26
Calibrate Buffer 2 (Two-Point Calibration only)
Immerse Sensor A in the second buffer solution.
and
are shown alternately on the upper display line. Use the
pH
Adjust
to set the displayed pH value to the temperature-corrected pH value of the chosen solution
(see the datasheet provided with the solution).
Press the
key to initiate calibration.
Note. To abort calibration, press the
complete – see previous page.
Measured pH value from the last successful calibration.
Measured sensor output in milivolts.
When the measured sensor output stabilizes, press the
is shown for 2 seconds to confirm that the calibration has been accepted,
Cal. Accepted
the display then advances automatically to the next frame.
The display then changes for 2 seconds to show the temperature-corrected buffer value in
the upper display line, then advances automatically to the next frame.
Calibration Message
See Table 4.1 for details of calibration messages.
Slope Value
% slope value.
A value between the programmed minimum % slope value (see
5.3) and 105% is displayed. If the value is outside these limits, check the electrode system.
pH check value.
Displayed as an additional indication of electrode system condition; 7pH is the optimum
value for glass electrodes and 0pH for Antimony electrodes.
Sensor B calibration (dual input analyzers only) is identical to Sensor A
Sensor Cal. B
calibration.
for Sensor B (dual input analyzers only) set to
SENSOR CAL.
Probe Type
set to
Enable Cals.
SECURITY CODE
Alter Sec. Code
CONFIG. DISPLAY
Alter Sec. Code
key again at any time before calibration is
(Section 5.3) – return to top of page.
No
not set to zero (Section 5.8) – see Section 5.1.
set to zero (Section 5.8) – see Section 5.2.
and
keys
key to accept the calibration.
– Section
Set Min Slope
or
Redox
ORP and