Configure Diagnostics - ABB AX416 User Manual

Single and dual input analyzers for ph/redox (orp)
Hide thumbs Also See for AX416:
Table of Contents

Advertisement

5.4 Configure Diagnostics

Note. The Configure Diagnostics function is applicable only if
Section 5.3.
-----
CONFIG. DIAGS
-----
Config. Sensor A
Off
-----
On
A: pH Glass
Off
-----
On
A: Ref. Checking
Config. Sensor A
Off
On
50
kOhm
A: Ref. Alarm
Config. Sensor A
Diff. Input
Configure Sensor A
Sensor B configuration (dual input analyzers only) is identical to Sensor A
Config. Sensor B
configuration.
pH Glass Check
Set to
to enable the impedance of the glass electrode circuit to be monitored in order
On
to provide low glass impedance and out-of-sample/broken cable warnings.
Note. This frame is displayed only if
– see Section 5.3.
Glass
Reference Electrode Check
Set to
to enable the condition of the reference electrode to be monitored in order to
On
provide:
– an indication of when sensor cleaning or replacement is required
– out-of-sample/broken cable warnings
Sensor B (dual input analyzers only) configuration is identical to Sensor A
Config. Sensor B
configuration.
See Section 5.5.
CONFIG. ALARMS
Reference Alarm
Set the impedence value above which the reference alarm is activated.
Sensor B (dual input analyzers only) configuration is identical to Sensor A
Config. Sensor B
configuration.
See Section 5.5.
CONFIG. ALARMS
5 PROGRAMMING...
for Sensor A and/or Sensor B is set to

is set to
A: Probe Type
pH
– see
Yes
is set to
A: Electrode
33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax466Ax468Ax436Ax460

Table of Contents