...7.3 Factory Settings
150. 0
Ohms
150. 0
Calib
A:T.Span (150R)
1000
Ohms
1000
Calib
A:T.Zero (1K0)
1500
Ohms
1500
Calib
A:T.Span (1K5)
2000
Ohms
2000
Calib
A:T.Zero (2K0)
5000
Ohms
5000
Calib
A:T.Span (5K0)
Cal. Sensor A
No
-----
Yes
A: Abort Cal.
Temperature Span (150R)
Set the temperature simulator to 150
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (1k0)
Set the temperature simulator to 1000
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Span (1k5)
Set the temperature simulator to 1500
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (2k0)
Set the temperature simulator to 2000
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Span (5k0)
Set the temperature simulator to 5000
The display returns automatically to
Abort Calibration
Select Yes or No
selected:
Yes
A:T.Zero (100R)
completion of
before
A:T.Zero (100R)
completion of
CAL. SENSOR A
after
Calibrate Sensor A
selected – calibration continues from the point at which the
No
pressed.
once a stable and valid value is recorded.
Cal. Sensor A
frame – calibration advances to
A:mV Span (+1V)
and continues.
frame – the display returns to the
A:mV Span (+1V)
page.
7 CALIBRATION...
key was
59