Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468

5.6 Configure Outputs

-----
CONFIG. OUTPUTS
-----
Config. Output 1
A-B
Temp.B
Sen.B
Temp.A
Sen.A
-----
AO1: Assign
4-20mA
0-20mA
0-10mA
-----
AO1: Range
pH
7. 0 0
pH
0. 0 0
AO1: Span Value
38
Configure Output 1
Output 2 configuration (and Outputs 3 and 4 if option board fitted and analog features
Config. Output 2
enabled - see section 7.3, page 58) is identical to Output 1 configuration.
Assign
Select the sensor and analog output required:
Sen.A
Sen.B
Temp.A
Temp.B
A-B
Note.
,
Sen.B
Temp.B
Range
Select the analog output current range for the selected output.
Span Value
pH (or
or
mV
Deg.C
display line. Use the
required span value.
Note. The minimum and maximum span values are determined by the
setting (see next page) plus the minimum differential, e.g. to set
first set
Zero Value
pH
Redox/ORP
Temperature
A-B
Note.
is applicable only to dual input analyzers.
A-B
Continued on next page.
AO1: Zero Value
-
pH/Redox (ORP)/mV for the selected sensor.
-
Temperature for the selected sensor.
-
Difference between the Sensor A and Sensor B readings.
and
are applicable only to dual input analyzers.
A-B
or
or
) and
Deg.F
A-B
Adjust
and
keys to adjust the displayed reading to the
to -2.00pH.
-
0.00 to 16.00 pH (minimum differential, 2.00pH)
-
-1100 to 1200 mV (minimum differential, 100mV)
-
Deg. C 0.0 to 150.0 (minimum differential, 10ºC)
Deg. F 32.0 to 302.0 (minimum differential,18ºF)
-
0.00 to 14.00 pH (minimum differential, 2.00pH)
5 PROGRAMMING
are shown alternately on the upper
Zero Value
to 0.00 pH,
Span Value
IM/AX4PH
Issue 9