Panel-Mount Analyzer Connections; Connections - ABB AX416 User Manual

Single and dual input analyzers for ph/redox (orp)
Hide thumbs Also See for AX416:
Table of Contents

Advertisement

...6.5 Panel-mount Analyzer Connections

6.5.2 Connections – Fig. 6.11
Terminal block A
85 to 265 V AC +
L
Line
or
N
Neutral
24 V AC
E
Connect supply earth (ground) to stud on case
A4 C
A5 NC
Relay 1
A6 NO
A7 C
A8 NC
Relay 2
A9 NO
A10 C
A11 NC
Relay 3 (see Note below)
A12 NO
A13 +
Analog Output 1
A14 —
A15 +
Analog Output 2
A16 —
Before making any electrical connections,
see Warnings on page 47
Note. Relay 3 can be configured to control the wash facility – see Section 5.4.
12
Power
to
Supplies
24V DC
Earth (Ground)
Stud on Case
(see Fig. 6.10)
Fig. 6.11 Connections, Panel-mount Analyzers
Terminal Block C
(Option Board)
C1
Not Used
Third Lead
C2
Not Used
C3
Used for optional
RS485 connections –
C4
refer to IM/PROBUS
C5
C6
Not Used
C7
C
C8
NC
Relay 4
C9
NO
C10 C
Third Lead
C11 NC
Relay 5
C12 NO
C13 +
Analog Output 3
C14 —
C15 +
Analog Output 4
C16 —
Refer to Section 6.6 for pH Sensor System
6 INSTALLATION...
Terminal Block B
Common
TC
Temperature Compensator

Connections

Common
TC
Temperature Compensator
Connections
Connection Details
B1
B2
B3
B4
B5
B6
B7
B8
B9
B10
B11
B12
B13
B14
B15
B16
53

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax466Ax468Ax436Ax460

Table of Contents