...4 SETUP
...4.1 Sensor Calibration

...4.1.4 Automatic, Single- and Two-Point Calibration (pH Only)

A: Buffer Method
set to
Auto 2-Pt
9. 1 8
pH
20. 0
Deg.C
A: Immerse Buf.2
128
mV
#### 100% ####
4. 0 1
pH
A: Cal. Buffer 2
A: Buffer Method
set to
Auto 1-Pt
-----
Passed
A: Calibration
100. 0
%
7. 0 0
pH
A: Slope & Check
Sensor Cal. A
24
Calibrate Buffer 2 (Two-Point Calibration only)
Immerse Sensor A in the second buffer solution.
Press the
key to initiate calibration.
Note. To abort calibration, press the
complete – see previous page.
The center display line shows the measured sensor output in millivolts.
As calibration procedes, a progress indicator appears in the lower display line. When the
measured sensor output stabilizes, the lower display line shows
The display then changes for 2 seconds to show the temperature-corrected buffer value in
the upper display line, then advances automatically to the next frame.
Calibration Message
See Table 4.1 for details of calibration messages.
Slope Value
% slope value.
A value between the programmed minimum % slope value (see
page – Section 5.3) and 105% is displayed. If the value is outside these
CONFIG. SENSORS
limits, check the electrode system.
pH check value.
Displayed as an additional indication of electrode system condition; 7pH is the optimum
value for glass electrodes and 0pH for Antimony electrodes.
Sensor B (dual input analyzers only) calibration is identical to Sensor A
Sensor Cal. B
calibration.
for Sensor B (dual input analyzers only) set to
SENSOR CAL.
Probe Type
set to
Enable Cals.
SECURITY CODE
Alter Sec. Code
CONFIG. DISPLAY
Alter Sec. Code
key again at any time before calibration is
(Section 5.3) – return to top of page.
No
not set to zero (Section 5.8) – see Section 5.1.
set to zero (Section 5.8) – see Section 5.2.
##### 100% #####
.
in the
Set Min Slope
or
Redox
ORP and