Recalibration
7.4 Recalibration
Note
This is the end of zero calibration. Proceed with zero, or span recalibration, or return to RUN.
7.4.2

Initial Zero

Perform an initial zero if necessary when calibration is out of range message is shown. During
this step the progress of the process will be shown on the display with a progress indicator in %.
Access P377 and enter EDIT mode
P377 Initial Zero
Enter 1 to start initial Zero
Press the following keys
Zero Calibration. Current Zero
Clear belt. Press ENTER to start
For example, the current zero.
Press the following key:
Initial Zero Calibration in progress
Current Reading:
For example, the zero count being calculated while calibration is in progress.
Calibration complete. Deviation
Press ENTER to accept value
For example: the deviation in % from previous zero. The new zero count if accepted. If unacceptable,
press the following key to restart:
74
E
0
530560
%
#####
0.00
551413
BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD