Zero Calibration - Siemens Milltronics BW500 Operating Instructions Manual

Integrators
Hide thumbs Also See for Milltronics BW500:
Table of Contents

Advertisement

6.4.3

Zero Calibration

Note
To obtain an accurate and successful calibration, ensure that the required criteria are met.
Refer to Calibration Criteria (Page 199).
Press the following key:
Zero Calibration: Current Zero
Clear belt. Press ENTER to Start
The current zero count
Press the following key:
Initial Zero Calibration. In progress
Current Reading:
The zero count being calculated while calibration is in progress
The duration of the Zero calibration is dependent on speed (P014), length (P016), and
revolutions (P360) of the belt.
Calibration Complete. Deviation
Press ENTER to accept value:
The deviation from previous zero. For an initial zero there is no previous zero; hence the deviation is
0. For example: the new zero count, if accepted
Press the following key:
Zero Calibration. Current Zero
Clear belt. Press ENTER to Start
Accepting the Zero returns to start of Zero. A new Zero can be performed, or continue to Span.
BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD
6.4 Load Cell Balancing
0
%
#####
0.00
551205
551205
Start Up
63