Access P359 In View Mode - Siemens Milltronics BW500 Operating Instructions Manual

Integrators
Hide thumbs Also See for Milltronics BW500:
Table of Contents

Advertisement

7.6.1

Access P359 in VIEW mode

P359 Factoring
Select 1-Weight, 2-Chain
Press the following keys:
Factoring Weight
Place weight and press ENTER.
For example: factor the test weight.
Press the following key:
Factoring Weight
Factoring in progress
The load reported while factoring is in progress.
Factoring Weight
Press ENTER to accept value
For example: the new value is accepted.
Press the following key:
P017 Test Load Weight:
Enter Test Load
For example: the current test load value.
Factoring is complete. Remove test weight and return to RUN mode if desired.
Note
If multispan function is used, the test load value is stored for the current multispan only.
BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD
Recalibration
7.6 Factoring
V
##.## kg/m
45.25
V
45.25
91

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Milltronics bw500/l

Table of Contents