P370 Zero Limit Deviation; P371 Auto Zero Initiation Upper Limit; P377 Initial Zero; P388 Initial Span - Siemens Milltronics BW500 Operating Instructions Manual

Integrators
Hide thumbs Also See for Milltronics BW500:
Table of Contents

Advertisement

12.5.10

P370 Zero Limit Deviation %

Sets the zero calibration deviation limit (±) from the last initial zero. If the accumulated deviation
of successive zero calibrations exceeds the limit, the zero calibration is aborted (f = 12.5).
Note
If the legal for trade certification switch has been set, the zero limit is ±2%.
Enter the maximum allowable % deviation, range is 1 to 12.5 %.
12.5.11

P371 Auto Zero Initiation Upper Limit

This parameter sets maximum load on the belt as a percentage of design load at which the
device will allow 'auto-zero' to run, min. 1.0%, max. 10.0%, default 2.0%.
When the certification switch is set this parameter will assume the default value (2.0%).
Note
Material will not be totalized during auto zero calibrations.
12.5.12

P377 Initial Zero

Resets the initial zero.
The initial zero is the reference zero to which all subsequent operator initiated zero calibrations
are compared in determining whether they have deviated beyond the zero limit (P370) (
Note
Refer to Initial Zero (Page 74) on for execution.
12.5.13

P388 Initial Span

Resets the initial span for the span selected, multispan 1 to 8 (P388-01 to -08)
The initial span is the reference to which all subsequent span calibrations are compared in
determining whether they have deviated beyond an accumulated ±12.5% of the initialspan (
1).
Note
Refer to Initial Span (Page 78) for execution.
BW500/L has one span entry possible, no indices will be presented.
1)
BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD
12.5 Online Calibration Options (P355 - P358)
Parameters
f
= 1).
.
1)
f
=
171