Inspecting the inner diaphragm
seal
Spanner
24 mm or 15/16 in
Spanner
(1.47 lbf.ft)
5 mm AF
2.6 Nm
(1.92 lbf.ft
)
Stylus and break stem assembly
1. Remove the stylus/break stem assembly
using the 5 mm AF spanner.
2. Use a 24 mm or 15/16 in spanner to remove
the probe's front cover. This will expose
the metal eyelid seal, spring and the inner
diaphragm seal. Remove the metal eyelid
and spring.
CAUTION:
!
Lose components may fall out.
www.armata-cnc.ru
3 Nm
Metal eyelid seal
Front cover
info@armata-cnc.ru
Inner diaphragm seal
Spring
3. Wash inside the probe, using clean coolant.
(DO NOT use sharp metal objects to clean
out debris).
4. Inspect the diaphragm seal for signs of
piercing or damage. In the event of damage,
return the probe to your supplier for repair, as
coolant entering the probe mechanism could
cause the probe to fail.
5. Refit the spring and metal eyelid (the spring's
largest diameter is against the metal eyelid).
6. Refit the remaining components.
+7(495)797-78-97
5.5