OTS installation guide
Changing the batteries
1
5.2
2
3
4
www.armata-cnc.ru
info@armata-cnc.ru
!
CAUTIONS:
Do not leave exhausted batteries in the probe.
When changing batteries, do not allow coolant or
debris to enter the battery compartment.
When changing batteries, check that the battery
polarity is correct.
Take care to avoid damaging the battery
compartment gasket.
Only use specified batteries (page 5.3).
Please dispose of exhausted batteries in
accordance with local regulations. Never dispose
of batteries in a fire.
NOTES:
Do not mix new and used batteries or battery
types as this will result in reduced life and
damage to the batteries.
Always ensure that the gasket and mating
surfaces are clean and free from dirt before
reassembly.
After removing old batteries wait more than 5
seconds before inserting new batteries.
If dead batteries are inadvertently inserted into
the probe then the LEDs will remain a constant
red.