Stylus level setting

The top surface of disc and square styli must be
set level, front to back and side to side.
Side to side level adjustment
3.7
1
2.5 mm AF
1.1 Nm
(0.81 lbf.ft)
Probe
module
Side to side level adjustment is obtained by
alternately adjusting grubscrews, which causes
the probe module to rotate and change the stylus
level setting.
When a level stylus surface is obtained, tighten
the grubscrews.