[Power On/Off + F4]⇒
Clearing only Active Memory
1. Turn off power (if it is currently on). Press <F4> and keep it pressed while you turn on power.
2. Press <Right Arrow> to access F4 Reset option list. Press <Down Arrow> to switch the
3. To restore operation, press <MENU> to access the installation menu. Create a new site
4. Re-select any Meter Facilities menu items (e.g., RS-232 setup parameters).
Clearing All Saved Memory
1. Turn off power (if it is currently on).
2. Press <F4> and keep it pressed while you turn on power. The prompt: [Clr Active Memory?
3. To access the F4 Reset option list press the <Right Arrow>. Press the <Down Arrow> to
4. To clear all Saved Memory press <ENT>.
5. Create a Site Setup before attempting to access other menu items.
6. To restore operation, press <MENU> to access the installation menu. Create a new site
7. Re-select desired Meter Facilities menu items (e.g., RS-232 setup parameters).
10.3

Test Facilities Graph Screen

When operating in the transit time mode the Test Facilities Graph Screen is an exceptional
diagnostic tool for troubleshooting problem applications or simply determining Receive signal
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01
The prompt: [Clr Active Memory? No] appears at the top of the screen.
option list to [Clr Active Memory? Yes]. Press <ENT> to clear all Active Site Data (but not
saved Site Setups).
setup or recall a stored site setup.
No] appears at the top of the screen. Press the <Down Arrow>. Note that the prompt
switches to [Clr Saved Data? No].
switch the option list to [Clr Saved Data? Yes].
CAUTION
Before proceeding further it is essential to understand that this function eliminates ALL
data stored in RAM. This means that all saved site setups including the site data of a flow-
calibrated site will be erased! In addition, the entire Datalogger file plus any custom factory
or user-created pipe or sensor tables will be eliminated. The impact of this is such that
we strongly recommend that you consult Technical Services before continuing with this
procedure. Be aware that you will have to create a new Site Setup, re-enter all site specific
parameters including pipe or sensor tables, plus all desired Meter Facilities menu entries.
setup and complete the installation procedure.
[Clr Active Memory?]⇒
⇑ ⇓
[Clr Saved Data?]⇒
Troubleshooting
10.3 Test Facilities Graph Screen
⇒No
⇒Yes
⇒No
⇒Yes
119