Connecting
5.2 Sensor Wiring
1. Perform all required menu steps up until the point where you respond to the [Install
2. Make a note of the Number Index displayed in the [Install Xdcr] menu. Check to ensure
①
Tension Screw
②
To SITRANS F 1010 Flowmeter
③
1012C Series Cable
④
Sensor Clamp
⑤
Upstream and Downstream Sensors
Figure 5-21
Reflect Mount with Model 1012TP Mounting Track (Side View)
3. Place the track rail assembly on the top surface of the pipe at the location where you have
50
Complete?] prompt.
that you have a matched set of sensors. They both should have the same S/N number but
marked with either an "A" or "B" (e.g., 100A and 100B).
Note
An index pin will be inserted into the track rail hole designated by the Number Index to
position one of the sensors (see
determined it will be mounted. Ensure that it is a smooth area without any raised spots or
seams.
⑨
in figure below).
⑥
Pipe
⑦
Ultrasonic Couplant
⑧
Number Index hole pin
⑨
1012TP Series Mounting Track
⑩
Flow direction
⑪
REF Hole Index Pin
Operating Instructions, 02/2010, A5E02951522A Revision 01
FUP1010 IP67 Portable Flowmeter