Table 7-9
RTD Calibrate
RTD Calibration by Entry Data
The [RTD Calibrate] menu allows you to adjust the intrinsic RTD reading to match an external
reference thermometer by directly entering its reading. Only perform this procedure while the
RTD under test is installed and currently measuring temperature.
To enter the current RTD temperature:
1. From the [RTD Calibrate] menu press <Right Arrow> to access the RTD option list.
2. Press <Right Arrow> to highlight the RTD you want to calibrate (RTD 1 or RTD 2).
3. Move the highlight to [Factory] or [User Cal] then press <ENT>.
4. This triggers the pop-up window:
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01
RTD Calibrate Menu Structure
RTD 1→
RTD 2→
Functions
7.10 RTD Calibration
Factory
User Cal
Factory
User Cal
95