Alarm, error, and system messages

8.1

Alarm Letter Codes and Descriptions

The following alarm codes appear on the main display of the flowmeter.
Table 8-1
Alarm Codes and Descriptions
Letter Codes
Alarm Code
SPACE
Spacing
EMPTY
Empty
HI/LO
Rate
FAULT
Fault
AER
Aeration
MEMRY
Memory
MAKUP
Makeup
I
Interface
P
Pig
Z
ZeroMatic
The displays shown below indicate where the Alarm Codes appear on the screen. Press <UP>
or <DOWN> Arrows to change screen views.
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01
Description
Transducer spacing may need adjustment
Pipe is empty
Flow above High setting or below Low setting
Three continuous seconds without new data update
Current aeration percentage exceeds the alarm set point
Last valid reading for a selected interval during Fault condition
In-Process Makeup occurred
The following alarm codes appear in the Datalogger status messages:
Liquid Vs exceeds interface alarm set point
Pig passage detected (optional)
ZeroMatic signal occurred