5.2 Sensor Wiring
Figure 5-13
Installation Procedure
1. Make a note of the Number Index displayed in the [Install Xdcr] menu. Check to ensure
2. Temporarily position one of the sensors on the pipe where you will be mounting it. Ensure
3. Prepare the area you marked by degreasing the surface and removing any grit, corrosion,
4. Attach the EZ Clamp to the sensor under the spring clip. The adjusting nut knob should be
42
Mylar Spacing Guide
that you have a matched set of sensors. They both should have the same S/N number but
marked with either an "A" or "B" (e.g., 100A and 100B).
that this is a smooth area without any raised areas (seams, etc.) With a pencil or chalk,
mark a generous area around all sides of the sensor. Remove the sensor.
rust, loose paint or surface irregularities with the abrasive pipe conditioning material
provided.
pointing up and on the side opposite of the spacer bar. Unscrew the knob until it's at the
stop.
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01