Functions
7.11 Reflexor
Selecting Sensor Mounting Location
Select a sensor mounting location that has a fully developed flow profile. Do not locate the
sensor so that sonic energy enters a region that is not representative of the flow velocities at
the measuring location. Use the figure below as a guide to select the proper mounting location.
Figure 7-1
98
Elbow
Expansion or Contraction
Sensor Mounting Examples
Valve
Orifice Plate
FUP1010 IP67 Portable Flowmeter