Siemens SITRANS FUP1010 Operating Instructions Manual page 146

Ultrasonic flowmeters 7me3510 portable flowmeter
Hide thumbs Also See for SITRANS FUP1010:
Table of Contents

Advertisement

Appendix
A.4 Site Setup For SITRANS F
● For site data or Waveshape data: filename.txt
● For Datalogger data: filename.csv
5. On PC type the proper command for the data desired (Logger, Site, or DP) and then [Enter].
5. The data now begins streaming on the HyperTerminal screen.
7. Wait for EOT (End Of Transmission) to be displayed.
8. Close the file by pointing to [Transfer], drag to Capture Text and click Stop button.
Closing the Terminal or HyperTerminal Program
You may now close the Terminal program. The file(s) you have downloaded are now saved
in the location you selected. You may now import the file you have saved into the appropriate
program (i.e., MS Word for site data, or MS Excel for Datalogger or Wave shape data for
graphing or analysis).
The Datalogger contains data that has its fields separated by commas. By using the file
extension ".csv" (comma separated values) suggested earlier, the data will import directly into
MS Excel without any further modification. For the wave shape data, the fields are separated
by spaces, therefore, it is best to save those files as .txt and then use the MS Excel Import
Wizard to select "Space Delimiters" for importation of the data.
Site data is downloaded in plain text and can be imported directly into MS Word.
To Clear Active Memory using the RS-232 Interface
1. Turn off power (if it is currently on). Turn power on. As soon as you apply power, immediately
2. Press the <Right Arrow> and then the <Down Arrow> to switch the option list to: [Clr Active
3. To restore operation, press <MENU> to access the Installation Menu. Create a new site
To Clear All Saved Memory using the RS-232 Interface
Before proceeding further, it is essential to understand that this function eliminates all data
stored in RAM. This means that all saved site setups, including the site data of a flow-
calibrated site will be erased! Also, the entire Datalogger file plus any custom factory or user-
created pipe or sensor tables will be eliminated. The impact of this is such that we strongly
recommend that you consult our technical service department before continuing with this
procedure. If you choose to continue, be aware that you will have to create a new site setup,
re-enter all site specific parameters including pipe or sensor tables, plus all desired Meter
Facilities entries.
144
type the @ character three times. The prompt: [Clr Active Memory? No] appears at the top
of the screen.
Memory? Yes] Press <Enter> to clear all Active Site Data (but not saved site setups).
setup or recall a stored site setup. Re-select any Meter Facilities items (e.g., RS-232 setup
parameters).
CAUTION
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01