Mounting Sensors
Two mounting configurations are available in the Reflexor Mode:
● Adjacent mounting locates the two sensors alongside each other using a single mounting
● In-Line mounting locates the two sensors axially along the pipe using two mounting chains
On inline mounted metal pipes the Receive Sensor is always located downstream of the
Transmit Sensor. On plastic pipes the Transmit Sensor is always located downstream of the
Receive Sensor.
Table 7-10
Sensor
C2 Sensor (A)
C2 Sensor (B)
In both cases, the cable entry will be on downstream side. Adjacent mounting will provide the
maximum sensitivity to flow. Refer to the appropriate sensor installation drawing for the
191N1S (Installation Drawing 191N1S-7).
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01
chain or strap.
or straps.
C2 Sensors
Flow Direction
Figure 7-2
Adjacent Sensor Mounting
Inline Metal and Plastic Pipe Cable Connections
Metal Pipe
DN
UP
Connect to DN on flowmeter
Connect to UP on flowmeter
Plastic Pipe
UP
DN
Functions
7.11 Reflexor
99