Troubleshooting

10.4 Site Setup Data

Key
<MENU>
<1>
<2>
<3>
<4>
<5>
<6>
<7>
<8>
<0>
<=>
<F1> and <.>
10.4
Site Setup Data
This menu provides data pertaining to sensor characteristics and operation. Some menu items
are for technical support interpretation only.
128
Command Line
MinDamp
MaxDamp
TnSet
ZCOSet
HiSet
LoSet
MaskSet
Hold Set
Description
Exits the Graph Screen and returns to
the main menu.
Minimum damping exponent control (+
or - to increase or decrease).
Maximum damping exponent control (+
or - to increase or decrease).
Transit time adjustment (use + or - to
move Tn marker).
Zero Crossover adjustment (use + or -
to move crossover marker).
Signal envelope threshold level (use +
or - to move threshold).
Signal envelope threshold level (use +
or - to move threshold).
Leading edge masking functions (use +
or - to alter number of samples masked).
Set this number higher if intermittent
mis-registration occurs.
Exits the command line.
Toggle graph between receive
waveform and envelope waveform.
Dumps the digitized waveform data over
the RS-232 port. You must first leave the
Graph Screen mode before invoking
this command.
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01