Siemens SITRANS FUP1010 Operating Instructions Manual page 46

Ultrasonic flowmeters 7me3510 portable flowmeter
Hide thumbs Also See for SITRANS FUP1010:
Table of Contents

Advertisement

Connecting
5.2 Sensor Wiring
7. While holding this sensor in place, bring the chain around the pipe and attach the closest
8. Tighten the adjusting nut until the chain is just snug around the pipe. Ensure that the chain
9. Tighten chain and ensure that the sensor does not move while tightening. Do not over
10.Attach the second sensor to the spacer bar at the index hole.
11.Do not apply any couplant to the second sensor. First attach the spacer bar assembly, with
12.Visually check that the spacer bar is axially aligned with the pipe.
13.Hold spacer bar assembly in this position and draw a pencil or chalk line along the back
14.Disassemble the spacer bar and second sensor from the mounted sensor. Remove the
44
cross-link onto the "S" hook of the EZ clamp.
is straight around the pipe and the sensor contacts the pipe at the white dot just under the
front label. Ensure that there is equal space on either side of the dot between the edges of
the sensor and pipe. Adjust if necessary.
tighten.
Note
In some cases the sensors may have two sets of holes for securing the spacer bar. When
using the EZ Clamp assembly use the lower set of holes when attaching the spacer bar.
the second sensor attached to it, to the mounted sensor by inserting an index screw through
the REF hole and into the lowest hole on the sensor. Tighten the screw.
edge of the sensor and a dot below the sensor label white dot. See "A" below.
second sensor from the spacer bar.
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01