12. Technical data

Measuring method:
Oscillometric and Korotkoff's principle of
blood pressure measurement
Display range:
0 – 300 mm Hg
Measuring range:
Systole (SYS): 50 – 250 mm Hg
Diastole (DIA):
Pulse:
40 – 160 beats / minute
Technical measuring precision: Cuff pressure: +/- 3 mm Hg
Pulse: +/- 5% of displayed pulse rate
Power supply:
4 x 1.5 V Mignon alkaline-manganese
(AA/LR06) batteries or optional
HARTMANN Tensoval mains adapter
Battery capacity:
Tensoval duo control:
> 1,300 measurements
Tensoval duo control large:
> 1,000 measurements
Inflation pressure:
approx. 190 mm Hg
Automatic switch-off function: 3 minutes after end of measurement
Cuff:
Standard cuff 22 – 32 cm
Large cuff 32 – 42 cm
Pressure release valve:
Electronically controlled linear valve
Memory capacity:
2 x 30 measurements and mean value
Operating conditions:
Ambient temperature: +10 °C to +40 °C
(+ 50 °F to + 104 °F)
Relative humidity:
15 – 90%
40 – 180 mm Hg
48
Storage / transport conditions: Ambient temperature: –20 °C to +50 °C
(– 4 °F to +122 °F)
Serial number:
See battery compartment
13. Power supply, disposal note
and safety information
13.1 Batteries, mains adapters
and disposal
The four high-quality batteries
included with the device guaran-
tee about 1,300 measurements.
Use only high-quality batteries
(see specification in Chapter 12
"Technical data"). If you are
using less efficient batteries we
can no longer guarantee 1,300
measurements.
Never mix old and new batteries
or batteries made by different
manufacturers.
Immediately remove exhausted
batteries.
If you do not intend to use the
device for a longer period of
time, you should remove the
batteries to prevent possible
leakage.
Please help protect the environ-
ment: Do not dispose of batte-
English
ries in household waste. Use
designated collection points or
municipal collection/recycling
centres to dispose of hazardous
household waste.
Information concerning the
disposal of electronic devices
(private households):
This symbol on products
and/or accompanying docu
ments means that spent elec-
tronic products must not be mixed
with regular household waste. Take
these products to designated col-
lection points where they are
accepted free of charge for proper
treatment, reclamation and recy-
cling. Proper disposal is in the
interests of environmental protec-
tion and prevents possible harmful
effects on people and the environ-
ment that may result from inappro-
priate handling of wastes. More
detailed information about your
nearest collection point is available
from your local council office.
49

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents