12. Technical data

Model:
Veroval
Type:
BPM25
Measuring method:
oscillometric
Display range:
0–300 mmHg
Measuring range:
Systolic (SYS): 50–280 mmHg
Diastolic (DIA): 30–200 mmHg
Pulse: 40–199 pulse beats per minute
Displaying of correct values outside the
measuring range cannot be guaranteed.
Display unit:
1 mmHg
Technical measure-
Cuff pressure: ± 3 mmHg
ment accuracy:
Pulse: ± 5% of the displayed pulse rate
Clinical measuring
Complies with the requirements of
accuracy:
EN 1060-4; Korotkoff validation method:
Phase I (SYS), Phase V (DIA)
Operating mode:
continuous mode
Nominal voltage:
DC 3V