Stroke Identification for Sample Container

S1:
Height identification of standard cup on sample
disk outer row (No. 1 position)
S2:
Height identification of cup on test tube/
same as for S1 (No. 2 position)
S3:
Height identification of standard cup on
sample disk inner row (W1 position) ................... STD/CONT
*1 Test tubes are settable only on the outer row. Model 902 allows height identification
for two kinds of sample containers.
(4)

Setting of Sample Containers at S.probe Adjustment

Container
Used
Cup-on-tube
Set
Position
S1
Standard cup
S2
Cup-on-tube
S3 (W1)
Standard cup
*2 After S.probe readjustment, check if the stroke down to the sample container has
changed.
Standard
Test Tube
Cup
Standard cup
Test tube
Standard cup/
Test tube
cup-on-tube
Standard cup
Standard cup
14 - 16
Sample containers
*2
Standard
Standard Cup/
Cup/Test
Cup-on-tube
Tube
Standard cup Standard cup
Test tube
Cup-on-tube
Standard cup Standard cup
*1
Test Tube/
Cup-on-tube
Test tube
Cup-on-tube
Standard cup