Hitachi 902 Service Manual page 44

Automatic analyzer
Table of Contents

Advertisement

Designation
No.
of Check
3
Calibration
(4) The input value for
parameter
check
(5) Relationship between
(6) STD CONC. must be
(7) When '99' is entered for
4
Volume
The following are checked
check
for measurable photometry
tests.
(1) The total liquid volume
(2) At least one of R1 to R3
(3) The reagent volume
(4) The minimum liquid
Details of Check
SPAN POINT must be
normal.
• When calibration
type is LINEAR (2 to
6-point) or LOGIT-
LOG (3P or 4P)
1) The above input
value must not
equal 0.
2) The above input
value must not
exceed that for
CALIB. POINTS.
calibration type and
calibration method
Refer to "output by each
calibration method and
check table" in 2.7.
(Note 2)
set in ascending order.
Check is made only for
necessary STD POS.
Note that isozyme STD
(3) and (4) are
excluded.
"=" is not regarded as in
ascending order.
POS. of STD (1),
CONC. must be 0.
up to the final
photometric point must
be 500 µL or less.
volumes must not be 0.
having a timing behind
the reaction time is not 0.
volume must be at least
250 µL.
Alarm
Display
Subdivision
CLB. PARAM?
1 to 37
(CH No.)
VOLUME ?
1 to 37
(CH No.)
2 - 32
(cont
d)
'
Remarks
NOTES:
2. Checked
against the
specification of
calibration test
selecting
.
information
When reagent
volume = 0, the
relevant reagent is
not used.