5.1
Functional Specifications of Model 902 ISE
Item
Measurable samples
Measuring temperature
Throughput
Measurement system
Data accuracy
Measuring concentration
range
Maintenance
Sampling
Printout
5. ISE
Serum, urine, blood
plasma
35 ± 2° C
Continuous measurement
Dilution to 1/50
Data reproducibility
Data accuracy
Serum
Urine
Rinsing operation
Reagent priming
Potential measurement
Upon simultaneous
request for photometry
and ISE
Upon independent request
for ISE
5 - 1
Specification
100 samples/hr max.
10 µL
Sample volume:
490 µL
Dilution volume:
Simultaneous reproducibility:
Serum
(N = 30)
Urine
Normal standard serum:
Na, Cl ±1.5 mmol/L
K
Na
:
80 to 180 mmol/L
K
:
1.5 to 10 mmol/L
Cl
:
60 to 120 mmol/L
Na, Cl :
10 to 250 mmol/L
K
:
1 to 100 mmol/L
Kind
:
Daily
Method : Automatic, pipetting from sample
cup
Internal standard, reference electrode
solution
Measurement of electromotive force of
Na/K/Cl in stand-by status
Priority given to ISE for calibrator, routine
sample and control sample
Data printed out immediately after
calculation
CV ≤ 1%
CV ≤ 2%
±0.1 mmol/L