10.7
DRV Board
(1)
Explanation of Circuit Board
The DRV board drives the stepping motor at a constant current/voltage and the solenoid
valve. The SSR control signal from the EMOT100 board is connected to the ACDIST
board via this board.
5 V and 24 V DC voltages are supplied to each unit in the instrument from the power unit
via this board.