(3) Test Selecting Information (from host to analyzer)
Send test selecting information corresponding to sample information.
1 2
Channel count
Initial measurement
NOTES:
1.
2.
Transfer sequence
3
4
5 ... 36 37
Photometry-assay channel

Details of Request for Each Channel

Function Character
" Ab "
" b "
" Db "
The above channel count ("bb0" to "b37") is the number of effective
channels from photometry-assay channel 1. If "bb3" is specified,
channels 1 to 3 are reflected on the analyzer with channels 4 to 37
ignored.
When there is at least one requested test, it is desirable to set "37."
Example:
1 2
3
b
b 3 1
1 1
Channel count
The analyzer side is obeyed if test request is made in a channel number
beyond the specified channel count. To avoid analysis for an
unnecessary test, set channel count "37" and request for only a test to be
specified.
Request for ISE is specifiable in the 37th channel.
(i) Specification of any other than "0" ... Select Na, K and Cl on the
screen.
(ii) Specification of "0" ... No request
Note that it is impossible to select a request for any of Na, K and Cl from
the host.
ISE channel
0:
No request
1:
Normal sample volume
2:
Unused
3:
Unused
4:
Volume determined by analyzer side
(left to analyzer)
4 5
36 37
1 .... 0
1
0 1
Not reflected on analyzer (analyzer side is followed)
Reflected on analyzer
16 - 40
Details of Request
(b: Space)