3)

ISE (electrolyte) type (1 character)

Table 16-21
ISE Type
'A'
Na, K
'B'
Na, K, Cl
4)

Data alarm for each channel (1 character)

A data alarm corresponding to each channel is transferred.
For details of data alarm, refer to the data alarm code list.
5)

ISE calibration data (72 characters)

This data area has eight data items; electromotive force of internal standard
solution, electromotive force of LOW solution, electromotive force of HIGH
solution, electromotive force of calibrator, slope level for display, concentration
of internal standard solution, concentration of calibrator and compensation
factor. Each data item is composed as shown below. Space remains blank
when there is no relevant data.
Data identification (2 bytes)
("bb" when there is no data)
Item
Electromotive force
of internal standard
solution
Electromotive force
of LOW solution
Electromotive force
of HIGH solution
Electromotive force
of calibrator
Slope level for
display
Concentration of
internal standard
solution
Concentration of
calibrator
Compensation factor
Test
(Unused)
Transfer sequence
Data alarm (1 byte)
Measured value (6 bytes)
Table 16-22
Data
Unit
Identification
" b1 "
mV
" b2 "
mV
" b3 "
mV
" b4 "
mV
" b5 "
mV
" b6 "
mEg/L 6 digits with sign and
" b7 "
mEg/L 6 digits with sign and
" b8 "
mEg/L 6 digits with sign and
16 - 37
Form
6 digits with sign and
decimal point
6 digits with sign and
decimal point
6 digits with sign and
decimal point
6 digits with sign and
decimal point
6 digits with sign and
decimal point
decimal point
decimal point
decimal point
Decimal Point
Position
1 digit
1 digit
1 digit
1 digit
1 digit
Same position as for
LOW solution on
ISE parameter
screen
Decimal point
position in calibrator
concentration on ISE
parameter screen