2.5

Retry Code Table

Code
Description
1
Alarm fuse blown
2
12 V for lamp
3
15 V for CPU rack
4
-15 V for CPU rack
14
Liquid in vacuum tank
15
Vacuum pressure abnormal
16
Waste solution tank
17
Incubation bath water inadequate
18
Distilled water short
19
Distilled water sensor abnormal
31
GPCNT1
32
GPCNT2
33
GPCNT3
34
GPCNT4
35
GPCNT5
36
GPCNT6
37
GPCNT7
38
GPCNT8
46
GMCNT1
47
GMCNT2
48
GMCNT3
49
50
ADC controller
71
Checksum error detected in operation
unit
72
Checksum error detected in analyzing
unit
102
103
104
105
106
107
Allowable
Retry Count
1/50 ms
1/4.5 sec
Check is not made when lamp is turned
off (during initialization, water exchange
or sleep).
1/50 ms
1/50 ms
1/4.5 sec
1/4.5 sec
Check is not made within 10 sec after
vacuum pump turns on.
1/4.5 sec
100/4.5 sec
Checked in the following statuses;
standby, operation, sampling stop.
1) 10/6 sec
Water supply
2) 50/6 sec
Warning
3) 150/6 sec
STOP
1/6 sec
Check is not made during initialization,
water exchange or wake-up.
10
Only at power-on
10
Only at power-on
10
Only at power-on
10
Only at power-on
10
Only at power-on
10
Only at power-on
10
Only at power-on
10
Only at power-on
10
Only at power-on
10
Only at power-on
10
Only at power-on
1
Only at power-on
1
1
2 - 64
Remarks