Hitachi 902 Service Manual page 23

Automatic analyzer
Table of Contents

Advertisement

Alarm
Alarm
Control
No.
1801 to
BARCODE2
1805
1931 to
REAGENT
1971
SHORT
2231 to
REAGENT
2271
LEVEL
2781 to
UN-
2830
ASSIGNED
2851,
TS OVER
2852
2861
PATIENT
SAMPLE
OVER
Sub-
Category
Level
code
54
1 to 5 WARNING An error has occurred in
1 to 40 WARNING •The volume of reagent to
57
59
1 to 40 WARNING Reagent volume is smaller
67
1
WARNING Because 400 samples are
68
1
WARNING Because 400 samples are
Description
communication with the
barcode reader. (Parity
error, framing error or
overrun error)
Sub-code indication
1 : Unassigned
2 : Unassigned
3 : Sample disk barcode
reader
4 : Unassigned
5 : Unassigned
be sipped from a reagent
bottle is inadequate.
•The reagent volume is a
total amount of one kind
of reagent for each test
item.
(1) Sub-code indicates
position No.
(2) Positions 37 and 38
are used for ISE.
(3) Position 40 is used for
HITERGENT.
(4) Position 39 is used for
HIALKALI.
than the remaining reagent
check value specified with
system parameter.
registered in routine sample
test selecting information,
new test selecting
information cannot be
registered from the host.
registered in routine sample
test selecting information, a
new routine sample cannot
be analyzed.
2 - 11
(cont'd)
Remedy
Check the
communication cable
of barcode reader.
Set new reagent.
Set reagent newly.
After completion of
analysis, register TS
again.
Same as above