5.4
ISE Measurement Sequence
For the dilution bath in ISE measurement, be sure to use two connected reaction cuvettes.
(1)
Sample Pipetting
Cuvette no.
n + 2
Cycle no.
Stop no.
(2)
Diluent Pipetting + Water + Stirring
Cuvette no.
n + 2
Cycle no.
Stop no.
(3)
Internal Standard Solution Pipetting + Water + Stirring
Cuvette no.
n + 2
Cycle no.
Stop no.
(4)
Sample Aspiration
Cuvette no.
n + 2
Cycle no.
Stop no.
S. probe
SMP.
n + 1
11
12
1
1
S. probe
DIL
SMP.
n + 1
11
12
3
3
S. probe
DIL
IS
SMP.
n + 1
12
13
3
3
ISE sipper
DIL
SMP.
IS
n + 1
17
18
1
1
Sample pipetting volume: Fixed at 10 µL
n
13
1
Diluent pipetting volume :
Water
n
13
3
Internal standard
solution pipetting volum
n
Water
14
3
Sample measurement
n
19
1
5 - 6
Fixed at 308 µL
Fixed at 182 µL
:
Total 490 µL
: Fixed at 318 µL
Fixed at 182 µL
:
Total 500 µL