Hitachi 902 Service Manual page 22

Automatic analyzer
Table of Contents

Advertisement

Alarm
Alarm
Control
No.
1441 to
ADC1?
1444
1451 to
ADC2?
1453
1461 to
ADC3?
1462
1471 to
BARCODE1
4506
Sub-
Category
Level
code
47
4
WARNING
48
1
WARNING Refer to alarm code 47-1.
48
2
WARNING Refer to alarm code 47-2.
48
3
WARNING Refer to alarm code 47-4.
49
1
WARNING Refer to alarm code 47-2.
49
2
WARNING
50
1 to 35 WARNING Data reception from the
Description
A/D count for 2 V is
abnormal.
A/D count for 6 V is
abnormal.
Normal count for 2 V
(7547 < count value <
8341)
Normal count for 6 V
(22460 < count value <
25023)
(This is checked at
measurement of
reference voltage.)
Reference voltage for
ISE: 2 V and 8 V
Normal count for 2 V
(7547 < count value <
8341)
Normal count for 8 V
(30184 < count value <
33364)
Normal offset value
(22811 < count value <
40547)
A/D count for 2 V is
abnormal.
A/D count for 6 V is
abnormal.
Normal count for 2 V
(12452 < count value <
13763)
Normal count for 6 V
(37356 < count value <
41288)
barcode reader has not
been completed before ID
reception time-out.
(Sub-code indicates the
position No. on disk.)
2 - 10
(cont'd)
Remedy
Same as above
Replace ISE AMP PC
board.
Same as above
Same as above
Replace EMIO100
PC board.
Same as above
Replace the label or
barcode reader.