Hitachi 902 Service Manual page 29

Automatic analyzer
Table of Contents

Advertisement

Alarm
Alarm
Control
No.
4511 to
CHEMISTRY
4548
PARAME-
TER?
4661 to
CALIBRA-
4698
TION
PARAME-
TER CALIB.
Sub-
Category
Level
code
98
1 to 36
WARNING (1) The relationship
(ch.)
99
1 to 36
WARNING (1) The relationship
(ch.)
2 - 17
Description
between assay code and
photometric point is
improper.
(2) The assigned
photometric point lags
behind the specified
reaction time.
(Operation is
impossible.)
between assay code and
calibration type is
improper.
(2) Necessary calibration
points for calibration
type are not input.
(3) Necessary standard
positions for calibration
are not input.
(4) The relationship
between calibration type
and calibration method
is improper.
(5) The standard
concentration values are
not set in ascending
order. (Except for STD
(3) and (4) for isozyme)
(6) Concentration value is
not zero when '99' is
entered for POS. of STD
(1).
1. The photometry assay
data in other than
manual mode is
checked.
2. Operation is
unallowable.
3. Check in (4) is made
with regard to
specifications of time-
out calibration, lot-to-lot
calibration, bottle-to-
bottle calibration and
test selecting
information.
(cont'd)
Remedy
Correct the
parameter.
Check parameter
and input it again.