Hitachi 902 Service Manual page 395

Automatic analyzer
Table of Contents

Advertisement

Inquiry Condition
[RUN] (highlighted) is specified for host
Operation
Instrument
Sample
communication on start condition screen, [NO] is
Mode
Mode
Name
specified for the transfer of analytical data alone on
the [communication parameter screen and [NO] is
specified for original Abs.
Simple
ID mode
Routine
Press the registration key on the
analysis
sample
screen by the same times as the
mode
sample count for inquiry.
Stat
When [RUN] (highlighted) is specified for stat sample
sample
test selection, inquiry is made while reading barcode
from the position specified for stat sample position
setting parameter.
Sample No.
Routine
Press the registration key on the
mode
sample
screen by the same times as the
sample count for inquiry.
Stat
Inquiry is not made.
sample
Routine
ID mode
Routine
Inquiry is made by reading
analysis
sample
barcodes sequentially from
mode
position No. 1.
Stat
When [RUN] (highlighted) is specified for stat sample
sample
test selection, inquiry is made while reading barcode
from the position specified for stat sample position
setting parameter.
Sample No.
Routine
Inquiry is made for the specified
mode
sample
number of samples sequentially
starting from the sample
specified in the analysis start
No. on start condition screen.
Stat
Inquiry is not made.
sample
Supplementary Explanation
1.
If no response is available from the host within 2 cycles (36 sec) after inquiry to the host, analysis is carried out according to the TS of main frame.
If the main frame has no TS request, the relevant sample will not be analyzed.
Table 16-12 Detailed Information about "SPE: Text for test selecting information inquiry (from analyzer to host)"
2 Characters
Function
Character
Sample
A
unregistered
Sample registered
D
Sample
N
unregistered
Sample registered
Sample
A
unregistered
Sample registered
D
Sample
N
unregistered
Sample registered
16 - 28
Sample Information
5 Characters
1 Character
3 Characters
Sample No.
Unassigned
Position No.
1 to 400
Space
1 to 35
1 to 50
Space
2 to 35
1 to 400
Space
1 to 35
Space
Space
1 to 35
Space
Space
2 to35
(position Nos.
read by
barcode
reader)
1 to 400
Space
1 to 35
1 to 400
TS: Test selecting information
13 Characters
15 Characters
ID No.
Unassigned
ID read with
Space
Even if a barcode read error occurs due
barcode reader
to an unregistered sample, inquiry is
made with position No. assigned.
Same as
Inquiry is not made when [NO] is
above
specified for full-time inquiry and TS
request lasts 1 channel or more.
Same as
Space
Same as above
above
ID registered
Space
Same as above
on screen
[handled as
comment]
ID read with
Space
Inquiry is not made when an
barcode reader
unregistered sample causes barcode
read error.
Same as
Inquiry is not made when [NO] is
above
specified for full-time inquiry and TS
request lasts 1 channel or more.
ID registered
Space
Same as above
on screen
When barcode reading is
successful, inquiry is
made via the read ID.
When unsuccessful,
inquiry is made via the ID
specified on screen
ID registered
Space
Same as above
on screen
[handled as
comment]
Remarks

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents