Calibrate Sensor

New
Reset
(Fig. 65)
Do you want to return to
Default calibration values?
Enter: Confirm
Esc: Cancel
(Fig. 66)
Set Reference and
wait for adjustment
Read value 26.9°C
Calib. value 27.0°C
(Fig. 67)
Calibration OK
Press any key to
continue
(Fig. 68)
49 - 60
Temperature Menu
Calibrate Sensor
This menu allows you to calibrate the Temperature probe
through the interface. By using the keys , you may choose
whether to perform a new calibration by selecting "New",
or to delete a previous one and reset the interface to the
default settings by selecting "Cancel" (Fig. 66). A calibration
should be performed when the interface is first connected to
the system.
ATTENTION:
standard calibration set by default, but it is advisable
to perform a calibration using a mercury thermometer
as reference.
In the event that the read temperature does not correspond
to other instruments in the aquarium, it is possible to align
the values by following the procedure below:
1) Immerse the sensor in your aquarium.
2) Enter in the menu "Calibrate sensor".
3) Using the keys  and  set the value read by the refer-
ence instrument next to "Calib. Value" (Fig. 67)
4) Wait 10 minutes in order to allow the sensor's read value
to stabilize. The read value may stabilize on a value that is
slightly different than that of the reference.
1/1
5) Once the 10 minutes have elapsed, press "Enter".
6) The controller will display the calibration result (Fig. 68);
rinse the probe and insert in aquarium.
Note: The calibration may be cancelled at any time by press-
ing "Esc". This will return the calibration parameters to those
of the last completed calibration.