  • ENGLISH, page 1
PG-gjennomsnitt = sum (4v
der:
v = topphastighet ved intervall n
N = antall intervaller i Riemann-summen
Qp/Qs
Oh, J.K., J.B. Seward, and A.J. Tajik. The Echo Manual. 3rd Edition, Boston: Philadelphia: Lippincott Williams
and Wilkins, (2007), s.70–72.
Qp/Qs = SV Qp-punkt/SV Qs-punkt = RVOT SV/LVOT SV
der:
RVOT SV = RVOT CSA* RVOT VTI = π/4 * RVOT-diameter
LVOT SV = LVOT CSA * LVOT VTI = π/4 * LVOT-diameter
Slagvolum (SV) doppler i ml
Oh, J.K., J.B. Seward, and A.J. Tajik. The Echo Manual. 3rd Edition, Boston: Philadelphia: Lippincott, Williams
og Wilkins, (2007), s.69–71.
SV = (CSA * VTI)
der:
CSA = åpningens tverrsnittareal (LVOT-areal)
VTI = tidsintegral for hastigheten til LVOT
Rengjøring og desinfisering
Denne delen inneholder instruksjoner for rengjøring og desinfisering av ultralydsystemet, transdusere og
tilbehør.
Følg anbefalingene fra FUJIFILM SonoSite ved rengjøring eller desinfisering av ultralydsystemet,
transduseren og tilbehøret. Følg anbefalingene for rengjøring i bruksanvisningen fra produsenten av det
eksterne utstyret når det skal rengjøres eller desinfiseres.
Rengjøring og desinfisering