3D MODEL EXPOSURE
48 Planmeca ProMax 3D s & 3D Classic with ProTouch
40 mm
Impression
on polystyrene disc
NOTE Ensure that you have selected the correct exposure
6. Place the model (impression or plaster cast) on the
polystyrene disc and ensure that:
•
The model faces the sensor and
•
The model is positioned so that the distance between
the cross formed by the volume center lights and the
front edge of the model is 40 mm.
Volume center
lights
7. In the Planmeca Romexis program, click 3D > Model
Capture. Refer to the Planmeca Romexis User's
Manual for details on Romexis functions.
8. In the window that appears, first select the material
you are exposing. Then click the option Start Capture.
values on the X-ray unit. The exposure values shown
here are only examples.
Sensor
User's Manual (3D)