3D MODEL EXPOSURE
(Part number 10030330)
Adjustable
adapter
in highest position
Patient support table
11.1.2 Selecting settings
44 Planmeca ProMax 3D s & 3D Classic with ProTouch
Polystyrene disc
Forward field
4. Remove all excess material from the top of the
calibration cup.
5. Gently pull the calibration pin out and ensure that the
surfaces of the hole formed by the calibration pin are
even (no air bubbles in inside walls).
6. Remove any patient supports attached to the X-ray
unit patient support table. Insert the polystyrene disc
provided (part number 10030330) into the adjustable
adapter.
7. Position the adjustable adapter so that it is in the
highest position.
Refer to section 9.1.3 "Adjusting adapter height" on
page 22 for details.
1. On the X-ray unit, select the program.
For impression material select
3D Models > Impression.
For plaster material select
3D Models > Plaster Cast.
The required exposure values depend on the material
and X-ray unit (X-ray tube and sensor) used. If you
need to adjust the preset exposure values (80 kV /
12.5 mA), proceed as described in section 9.2.4
"Adjusting exposure values for current exposure" on
page 26.
2. Select the forward field on the touch screen. The
positioning lights (volume center lights, volume bottom
light and incisor light) come on. The volume center
lights cross in the middle of the image volume.