Function
Toggles between 1 and 2. Used in CyDye DNA application only to set the number of dyes
used in the measurement.
Numeric entry. This option is only used if the counter ion type in the T
Other.
Used in User Access to set a password for new users.
Used in the nucleic acid applications, Protein A280 and Protein UV. This is the pathlength of
the cuvette used in the measurement. Options are for 10, 5, 2 and 1 mm.
Toggles between yes and no. Used in T
phosphorylated or not.
Numeric entry to 3 decimal places. Used in T
in pmole/mL.
Used in all applications to set the desired print location. Only available printers are
displayed.
Toggles between On and Off, used in Equation Editor only. With Prompt on the
measurement will proceed as follows: measure wavelength 1, prompt for sample, measure
wavelength 2, prompt for sample etc.
Used in all standard curve applications. This is the number of times a standard
measurement is repeated before the mean of these values is plotted on the standard curve.
Options are off (1 measurement), 2 or 3.
Used in wavescan measurements, this determines how many samples will be overlaid on
the graph. Options are off, 2 to 8.
Used in kinetic measurements, this sets the number of samples that will be measured
during the method. Options are 1, 2 or 3.
Used in all applications to set the desired save location. Only available locations are
displayed.
Toggles between on and off. Used in DNA/RNA/Oligo measurements this sets whether, as a
further analytical tool, the instrument will also perform a scan over the wavelength range
220-320 nm.
Used in wavescan measurements, this determines the speed at which the scan is
performed. Options are Slow, Medium, Fast or Integration. Note that when integration is
selected the user will be prompted to set the integration time (see box above).
Used in User Interface in Settings to set the time before the screen-saver will be
displayed.
Used by the Default Administrator in User Access to set if user login will be displayed or not.
Used in all standard curve applications. This is the number of standards that will be used to
create the standard curve; options are from 1 to 9.
Numeric entry to 2 decimal places. Used in all standard curve applications this is the
concentration of the standard.
Used in wavescan measurements, this determines the interval at which wavescan
measurements are made (in nm). Options are for 0.1, 0.2, 0.5 or 1 nm. Smaller intervals give
greater detail but lead to longer scan times.
calculation to set if the sample to be measured is
m
calculation to sets the primer concentration
m
76
calculation is set to
m
75