There are no bio-hazardous materials within the unit; however this unit could be used with bio-hazardous samples.
Before using the instrument the customer should have in place decontamination procedures designed to protect
laboratory workers from occupationally acquired infections. The sample chamber cell holders are removable and
may be decontaminated using the appropriate disinfectant for the bio-hazard in question, rinsed with distilled
water and then allowed to dry. The sample chamber and exterior may be wiped with a suitable disinfectant
cleaning wipe.
• D econtamination. Equipment returned for repair should include an appropriate decontamination certificate.
• I t is the responsibility of the customer to ensure that the user of the equipment is provided with a safe working
environment.
• A ny chemicals used in Analyses should be used, stored and disposed of in accordance with manufacturer's
guidelines and local safety regulations.
• T oxic Fumes. Efficient laboratory ventilation must be provided when working with volatile solvents or toxic
substances.
• W aste disposal. Disposal of some solvents and chemicals may be classed as hazardous waste and must be
dealt with in accordance with local regulatory practice.
• P ersonal protective equipment. This is not required to operate the unit but the samples measured may require
PPE. A local risk assessment should be carried out.
This equipment may be connected and controlled from a PC. To preserve the integrity of the measuring equipment
it is essential that the attached PC itself conforms to basic safety and EMC standards and is set up in accordance
with the manufacturers' instructions. If in doubt consult the information that came with your PC. In common with
all computer operation the following safety precautions are advised.
• T o reduce the chance of eye strain, set up the PC display with the correct viewing position, free from glare and
with appropriate brightness and contrast settings.
• T o reduce the chance of cross contamination from biological samples, use appropriate personnel protection
measures and disinfectant wipes on keyboard and mouse.
Care must be taken when handling all heated accessories.
1.2. Unpacking & Installation
• These instruments weigh approximately 18.5 kg. Follow your local regulations for
safe handling and lifting of this equipment.
• Inspect the instrument for any signs of damage caused during transit. If any
damage is discovered. Do not use the instrument and report the problem to your
supplier.
• The instrument must be placed on a stable, level bench or table capable of taking
its weight with sufficient space around the instrument for ventilation to circulate
freely.
• Ensure your proposed installation site conforms to the environmental conditions for
safe operation.
• Indoor use.
• 5 to 40°C.
• Maximum relative humidity 90% up to 31°C decreasing linearly to 50% at 40°C.
• Extremes of temperature may require re-calibration of the unit for optimum
performance.
• If the instrument has been stored in a cold environment then it should be allowed
to come to thermal equilibrium for 2 to 3 hours before operation so that start up
calibration is not compromised by condensation.
• The equipment must be connected to the local supply outlet using the provided
power cables. It can be operated from 90 to 264 V~ 50 or 60 Hz.
• Replace power inlet fuses only with the same type and rating as follows:
5061-046 Rev1
5
Need help?
Do you have a question about the ultrospec 7000 and is the answer not in the manual?