Anritsu MS2024A/MS2026A User Manual page 174

Vna master vector network analyzer 2 mhz to 6 ghz spectrum analyzer 9 khz to 7.1 ghz
Table of Contents

Advertisement

4-33 Trace
Trace A Operations
Allows the user to select from the operations that can be applied to Trace A.
Normal -> A
The trace displays current data for the current sweep.
Max Hold -> A
The trace shows the cumulative maximum value of each display point over many
sweeps.
Min Hold -> A
The trace shows the cumulative minimum value of each display point over many
sweeps.
Average -> A
The trace shows an exponential average of a number of traces, determined by the
# of Averages key, below.
# of Averages
Sets the number of traces that are used for the calculation of the average display
value from 1 to 65535.
Back
Returns to the previous menu.
Trace B Operations
Allows the user to select from the operations that can be applied to Trace A.
A -> B
Copies the contents of Trace A into Trace B. Doing so overwrites the previous
contents of Trace B.
B <--> C
Swaps the contents of Traces B and C.
Max Hold -> B
Trace B shows the cumulative maximum value of each display point over many
sweeps.
Min Hold -> B
The trace shows the cumulative minimum value of each display point over many
sweeps.
Back
Returns to the previous menu.
Trace C Operations
Allows the user to select from the operations that can be applied to Trace C.
A -> C
Copies the contents of Trace A into Trace C. Doing so overwrites the previous
contents of Trace C.
B <--> C
Swaps the contents of Traces B and C.
4-62
Chapter 4 — Spectrum Analyzer
MS20xxA VNA Master UG 10580-00166 Rev. C

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MS2024A/MS2026A and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Ms2034a/ms2036a

Table of Contents